RAYSEARCH LABORATORIES AB: Medical Device Recall in 2017 - (Recall #: Z-1815-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Radiation Therapy Treatment Planning System, Model 4.5, 4.7, 5.0 RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs

Product Classification:

Class II

Date Initiated: January 18, 2017
Date Posted: April 19, 2017
Recall Number: Z-1815-2017
Event ID: 76342
Reason for Recall:

An error in NVIDIA GPU (Graphics Processing Unit) card drivers can occur for certain software programs. In a non-standard Citrix environment where the GPU settings are configured with a virtual Citrix graphics board (graphics card) an erroneous calculation may be allowed.

Status: Terminated
Product Quantity: 741 units
Code Information:

4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11, 5.0.2.35

Distribution Pattern:

Nationwide Distribution to AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, LA, MN, MD, MA, MI, MS, MT, NJ, NY, NC, NV, OH, OK, OR, PA, RI, TN, TX, VA, VT, VI, WA, WI, and Washington D.C.

Voluntary or Mandated:

Voluntary: Firm initiated