RAYSEARCH LABORATORIES AB: Medical Device Recall in 2020 - (Recall #: Z-0736-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

RayStation stand-alone software treatment planning system, Model Numbers 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0 - Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Product Classification:

Class II

Date Initiated: November 18, 2019
Date Posted: January 1, 2020
Recall Number: Z-0736-2020
Event ID: 84508
Reason for Recall:

Editing a static arc beam may unexpectedly set the MU of the beam to its initial default value of 200 MU. If the bug is triggered, the dose is invalidated and needs to be recalculated before the plan can be approved or exported.

Status: Terminated
Product Quantity: 956 Devices/Licences
Code Information:

Software Version Numbers: 4.0.0.14 , 4.0.1.4 , 4.0.2.9 , 4.0.3.4 , 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.9.0.42 , 5.0.0.37 , 5.0.1.11 5.0.2.35 , 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , or 9.0.0.113

Distribution Pattern:

US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, and WI.

Voluntary or Mandated:

Voluntary: Firm initiated