RAYSEARCH LABORATORIES AB: Medical Device Recall in 2020 - (Recall #: Z-2352-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
RayStation 6 Service Pack 1, UDI #s 07350002010037, 07350002010082 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Class II
Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
UDI #s 07350002010037, 07350002010082 Build number: 6.1.1.2
US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
Voluntary: Firm initiated