Product Description:

TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52, 10.1.0.613, 11.0.0.951 or 11.0.1.29.

Product Classification:

Class II

Date Initiated: September 9, 2021
Date Posted: April 6, 2022
Recall Number: Z-0848-2022
Event ID: 89788
Reason for Recall:

If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning

Status: Terminated
Product Quantity: 51 (US) Licenses
Code Information:

Name (build number) UDI-DI: RayStation 6/RayPlan 2 Service Pack 1 (6.1.0.26) 07350002010037 RayStation 6/RayPlan 2 Service Pack 1 v.6.1.1 (6.1.1.2) 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) 07350002010242 RayStation/RayPlan 7 (7.0.0.19) 07350002010068 RayStation/RayPlan 8A (8.0.0.61) 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) 07350002010136 RayStation/RayPlan 8B (8.1.0.47) 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) 07350002010235 RayStation 9A (9.0.0.113) 07350002010174 RayStation 9B (9.1.0.933) 07350002010266 RayStation Service Pack 1 (9.2.0.483) 07350002010297 RayStation 10A (10.0.0.1154) 07350002010303 RayStation 10A Service Pack 1 (10.0.1.52) 07350002010365 RayStation 10B (10.1.0.613) 07350002010310 RayStation 11A (11.0.0.951) 07350002010389 RayStation 11A Service Pack 1 (11.0.1.29) 07350002010433

Distribution Pattern:

US Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS, NC, OK, OR, SC, TN, WA, WI.

Voluntary or Mandated:

Voluntary: Firm initiated