RAYSEARCH LABORATORIES AB: Medical Device Recall in 2023 - (Recall #: Z-0206-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care Product name (build number): (1)RayCare 5B SP1 (5.1.1.60246); (2)RayCare 5B SP3 (5.1.3.60023); (3) RayCare 6A (6.0.0.60553)

Product Classification:

Class II

Date Initiated: September 12, 2023
Date Posted: November 8, 2023
Recall Number: Z-0206-2024
Event ID: 93139
Reason for Recall:

An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.

Status: Ongoing
Product Quantity: 2 sytems
Code Information:

UDI-DI: (1)RayCare 5B SP1 (5.1.1.60246) 0735000201053220220316; (2)RayCare 5B SP3 (5.1.3.60023) 0735000201069320221027; (3) RayCare 6A (6.0.0.60553) 0735000201056320220617

Distribution Pattern:

MD, TN

Voluntary or Mandated:

Voluntary: Firm initiated