RAYSEARCH LABORATORIES AB: Medical Device Recall in 2026 - (Recall #: Z-1102-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System
Class II
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
UDI: 0735000201031020201216. GTIN: 07350002010310. Serial Numbers: 10.1.0.613. Software Revision: RayStation 10B. Expiration Date: 2027-02-01.
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Voluntary: Firm initiated