RAYSEARCH LABORATORIES AB: Medical Device Recall in 2026 - (Recall #: Z-1103-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatment Planning System

Product Classification:

Class II

Date Initiated: November 28, 2025
Date Posted: January 28, 2026
Recall Number: Z-1103-2026
Event ID: 98117
Reason for Recall:

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Status: Ongoing
Product Quantity: 10 units
Code Information:

Lot Code: UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916, 35000201063120220616. GTIN: 07350002010389, 07350002010433, 07350002010440, 07350002010631. Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116, 11.0.4.15. Software Revision: RayStation 11A, 11A SP1, 11A SP2, 11A SP3. Expiration Date: 2027-06-22.

Distribution Pattern:

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Voluntary or Mandated:

Voluntary: Firm initiated