Rebiscan, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0168-2023)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Blinq Pediatric Vision Scanner-Pediatric Vision Scanner-intended for automated detection of misalignment of the visual axes using polarized light. Model Number BQ830
Class II
Unable to charge the battery pack- battery packs will allow the batteries to be charged if the voltage of any cell is below 2.5V. When this occurs, the battery pack will not accept the charge, could vent and possibly catch fire
UDI: 008648160004 Serial Numbers: B-00001-V thru B-00105-V B-00021 B-00022 B-00036 B-00052 B-00054 thru B-00060 B-00062 thru B-00065 B-00070 B-00071 B-00075 thru B-00077 B-00079 B-00082 thru B-00098 B-00100 thru B-00104 B-00106 B-00109 B-00111 thru B-00113
Worldwide distribution - US Nationwide and the countries of New Zealand, Belgium.
Voluntary: Firm initiated