Product Description:

Thermo SCIENTIFIC, Sensititre STP6F, Gram Negative IVD AST

Product Classification:

Class I

Date Initiated: October 20, 2022
Date Posted: December 14, 2022
Recall Number: Z-0334-2023
Event ID: 91104
Reason for Recall:

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Status: Terminated
Product Quantity: 2656 units
Code Information:

Lot Numbers: B2314, B2174, B2036, B1346, B1183, B1115, B1027

Distribution Pattern:

US Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated