Remel, Inc: Medical Device Recall in 2025 - (Recall #: Z-1395-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Zebra Thermal Barcode Printer GX430t, Model no. YSW 1000 or YSW 2000, may have been sold as an optional accessory for Sensititre Aris HIQ system YV4000-VZ

Product Classification:

Class II

Date Initiated: February 12, 2025
Date Posted: March 26, 2025
Recall Number: Z-1395-2025
Event ID: 96278
Reason for Recall:

Power supply unit for Zebra Printers used with the device can potentially overheat or cause a fire hazard.

Status: Ongoing
Product Quantity: 41
Code Information:

UDI-DI 05032384545872. Detailed list of affected Zebra printer models: https://www.zebra.com/us/en/power-supply-recall.html Please check the manufacturing label found either on the back or bottom of your printer. If the label lists Zebra Technologies as the manufacturer, please use the validation form on the Zebra website to fully validate if your power supply is affected: https://connect.zebra.com/safeguard

Distribution Pattern:

Domestic distribution to AL, AZ, CA, FL, GA, IN, KS, KY, MD, MI, NE, NY, OH, PA, TN, TX, WI. International distribution to Canada and the Bahamas.

Voluntary or Mandated:

Voluntary: Firm initiated