Remington Medical Inc.: Medical Device Recall in 2014 - (Recall #: Z-1441-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
FAS-LOC 6 Foot Disposable Extension Cable with Safe Connect/Screw Down. Electrical extension cable intended to transmit signal from, or power or excitation signal to patient connected electrodes. The cable is a sterile, disposable device.
Product Classification:
Class II
Date Initiated: March 5, 2014
Date Posted: April 16, 2014
Recall Number: Z-1441-2014
Event ID: 67752
Reason for Recall:
Remington Medical discovered an error on the FL-601-97 IFU; part number IFU-RM-0154. Specifically, the polarity for the black chuck in the picture is wrong. It shows (+), but should show (-).
Status: Terminated
Product Quantity: 41,499 cables
Code Information:
Lot numbers 131901, 132262, 132263, 132533, 132693, 132694, 132833 132835, 132986, and 133171.
Distribution Pattern:
USA (nationwide) Distribution including Canada
Voluntary or Mandated:
Voluntary: Firm initiated