Remote Diagnostic Technologies Ltd.: Medical Device Recall in 2025 - (Recall #: Z-2175-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R

Product Classification:

Class II

Date Initiated: May 30, 2025
Date Posted: August 6, 2025
Recall Number: Z-2175-2025
Event ID: 97197
Reason for Recall:

Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.

Status: Ongoing
Product Quantity: 55
Code Information:

REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28

Distribution Pattern:

Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.

Voluntary or Mandated:

Voluntary: Firm initiated