RESPIRATORY THERAPEUTICS GROUP LLC: Medical Device Recall in 2020 - (Recall #: Z-1443-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.
Class II
It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.
Lot KS1901013: 20019/03/31 - 06/192019 Lot KS1902002: 2019/06/21 - 07/31/2019
US Nationwide distribution in the states of CA, NJ, NY. OUS: None
Voluntary: Firm initiated