Respironics California, LLC: Medical Device Recall in 2020 - (Recall #: Z-1155-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Respironics V680 Ventilator, Service # 850011, (OUS Only)

Product Classification:

Class II

Date Initiated: September 11, 2019
Date Posted: February 19, 2020
Recall Number: Z-1155-2020
Event ID: 83980
Reason for Recall:

Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.

Status: Ongoing
Product Quantity: 190
Code Information:

Serial Numbers: 100255874, 100255957, 100257487, 100257488, 100257490, 100257491, 100257493, 100257494, 100257495, 100257496, 100257497, 100257498, 100257499, 100257500, 100257501, 100258060, 100258061, 100258063, 100258064, 100258065, 100258066, 100258067, 100258068, 100260220, 100260221, 100260222, 100260223, 100260224, 100260227, 100260228, 100260229, 100260616, 100260617, 100260618, 100260619, 100260620, 100260621, 100260622, 100260623, 100260624, 100260625, 100260626, 100260627, 100260628, 100260629, 100260630, 100260631, 100260632, 100260633, 100260634, 100260635, 100261843, 100261844, 100261845, 100261846, 100261847, 100261848, 100261849, 100261850, 100261851, 100261852, 100261853, 100261854, 100261855, 100261856, 100261857, 100261900, 100261901, 100261902, 100263992, 100263993, 100263994, 100263995, 100263996, 100263997, 100263998, 100263999, 100264020, 100264021, 100264022, 100264023, 100264024, 100264025, 100264026, 100264279, 100264300, 100264301, 100264302, 100264303, 100264561, 100264562, 100264563, 100264564, 100264565, 100264566, 100264567, 100264568, 100264569, 100264570, 100264571, 100264572, 100264573, 100264574, 100264575, 100264576, 100264577, 100264578, 100264579, 100264780, 100264781, 100264782, 100264783, 100264784, 100264785, 100265624, 100265625, 100265626, 100265627, 100265628, 100265629, 100265630, 100265631, 100265632, 100265633, 100265634, 100265635, 100265636, 100265637, 100265638, 100265639, 100265640, 100265641, 100265642, 201034917, 201034918, 201034919, 201034945, 201034946, 201034947, 201034949, 201034950, 201034951, 201034952, 201034953, 201034954, 201035015, 201035016, 201035017, 201035018, 201035019, 201035080, 201035081, 201035082, 201035083, 201035084, 201035085, 201035197, 201035198, 201035199, 201035221, 201035222, 201035223, 201035224, 201035225, 201035226, 201035320, 201035321, 201035322, 201035323, 201035324, 201035325, 201035326, 201035328, 201035329, 201035330, 201035331, 201035332, 201035333, 201035334, 201035335, 201035336, 201035337, 201035338, 201035339, 201035360, 201035361, 201035362, 201035363, 201035364, 201035365

Distribution Pattern:

US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam

Voluntary or Mandated:

Voluntary: Firm initiated