Roche Diagnostics Operations, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0168-2014)
See the recall detail below. You can also see other recalls from the same firm in 2013.
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.
Class II
Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof
software version 9864.C2
Nationwide Distribution and Puerto Rico
Voluntary: Firm initiated