Roche Diagnostics Operations, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1581-2013)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2013.
Data Source: FDA.
Product Description:
CoaguChek XS System (PST) The CoaguChek XS PT test strips are part of the CoaguChek XS System. The CoaguChek XS System measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin¿ or warfarin.
Product Classification:
Class II
Date Initiated: May 16, 2013
Date Posted: July 3, 2013
Recall Number: Z-1581-2013
Event ID: 65246
Reason for Recall:
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
Status: Terminated
Product Quantity: 204,205 meters total
Code Information:
Model/Catalog/Part Number: 04837738001
Distribution Pattern:
Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated