Roche Diagnostics Operations, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1927-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.
Class II
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Catalog numbers 04714423190; 04711220190; 05230438190; 05230446190.
US (nationwide) including PR.
Voluntary: Firm initiated