Roche Diagnostics Operations, Inc.: Medical Device Recall in 2016 - (Recall #: Z-0099-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
URISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a computer-controlled stand-alone sorting system for sorting barcoded samples. The system has an online connection to the host.
Product Classification:
Class II
Date Initiated: July 15, 2016
Date Posted: November 2, 2016
Recall Number: Z-0099-2017
Event ID: 75186
Reason for Recall:
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Status: Terminated
Product Quantity: 9014 in total
Code Information:
Not applicable
Distribution Pattern:
US Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated