Roche Diagnostics Operations, Inc.: Medical Device Recall in 2017 - (Recall #: Z-1092-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High Sensitive Immunoturbidimetric assay is for the in-vitro quantitative determination of C-reactive protein (CRP) in human serum and plasma on Roche automated clinical chemistry analyzers. Highly sensitive measurement of CRP is of use for the detection and evaluation of inflammatory disorders, and associated diseases, infection and tissue injury. Measurement of CRP may also be used as an aid in the assessment of the risk of future coronary heart disease. When used as an adjunct to other laboratory evaluation methods of acute coronary syndromes, it may also be an additional independent indicator of recurrent event prognosis in patients with stable coronary disease or acute coronary syndrome

Product Classification:

Class II

Date Initiated: February 26, 2016
Date Posted: February 1, 2017
Recall Number: Z-1092-2017
Event ID: 75711
Reason for Recall:

Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 11972855216) shows an under-recovery of up to -25% at concentrations below 5 mg/L on the MODULAR ANALYTICS P module. this issue can lead to erroneous low CRPHS results. CRPHS is used as a marker for the prediction of cardiovascular risk. An erroneous low result could lead to an incorrect cardiovascular risk assessment and subsequent delay of further examination.

Status: Terminated
Product Quantity: 3204
Code Information:

Catalog number: 11972855216, lot number 60445001

Distribution Pattern:

Nationwide Distribution including AL, AZ, CA, CT, FL, GA, KS, MA, MO, NJ, NY, PA, and WA.

Voluntary or Mandated:

Voluntary: Firm initiated