Roche Diagnostics Operations, Inc.: Medical Device Recall in 2020 - (Recall #: Z-1158-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
cobas infinity core, Material No. 7154003001. The device is a calculator/data processing module, for clinical use.
Product Classification:
Class II
Date Initiated: January 7, 2020
Date Posted: February 19, 2020
Recall Number: Z-1158-2020
Event ID: 84697
Reason for Recall:
Alarm flags that have a "<" symbol in front of them are not displayed on the cobas infinity Validation screen. This issue affects only customers that are doing manual validation on the cobas infinity Validation screen.
Status: Terminated
Product Quantity: 26 accounts
Code Information:
All cobas infinity laboratory solution software versions are affected.
Distribution Pattern:
Distributed to customers in AK, AL, CA, GA, HI, IA, IL, KY, MA, MI, MN, MO, MS, MT, NJ, NY, OR, SC, and TN.
Voluntary or Mandated:
Voluntary: Firm initiated