Roche Diagnostics Operations, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0266-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is comprised of the cobas u 601 urine analyzer and the cobas u pack. The cobas u 601 urine analyzer when used with the cobas u pack is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, color and erythrocytes, as well as clarity.

Product Classification:

Class II

Date Initiated: October 18, 2021
Date Posted: November 24, 2021
Recall Number: Z-0266-2022
Event ID: 88879
Reason for Recall:

A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinine levels up to 9,000 mg/L.

Status: Terminated
Product Quantity: 636
Code Information:

UDI 04015630932252

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated