Roche Diagnostics Operations, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0804-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
CoaguChek XS PST Meters- IVD (Patient Self-Testing) measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin or warfarin. Catalog 04837738001
Product Classification:
Class II
Date Initiated: December 2, 2020
Date Posted: January 13, 2021
Recall Number: Z-0804-2021
Event ID: 86951
Reason for Recall:
Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
Status: Terminated
Product Quantity: 716,880 units
Code Information:
All serial numbers
Distribution Pattern:
US Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated