Roche Molecular Systems, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0762-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

COBAS TaqMan Mycobacteria tuberculosis (MTB) Test FOR IN VITRO DIAGNOSTIC USE.

Product Classification:

Class II

Date Initiated: November 28, 2012
Date Posted: February 6, 2013
Recall Number: Z-0762-2013
Event ID: 63997
Reason for Recall:

The COBAS TaqMan MTB Test Definition Files (TDF) v3.0, for both CE-IVD and JPN-IVD tests, being used with AMPLILINK software version 3.3 series and COBAS TaqMan MTB Test Kits, both CE-IVD and JPN-IVD, have an incorrect Target Elbow Max cutoff value of 48 creating the potential for false negative results to be generated for patient samples with Target Elbows in the range of 48 to 50.

Status: Terminated
Product Quantity: 335 CE-IVD TDFs; 48 JPN-IVD TDFs
Code Information:

Material Number: CE-IVD: 04803531190 Kit, 05574455190 TDF; JPN-IVD: 04653360190 Kit, 05574447190 TDF. Lot/Serial Number: CE-IVD: All Kits, JPN-IVD: All Kits.

Distribution Pattern:

Worldwide Distribution - Foreign (22 EU countries - Austria, Belgium, Bulgaria, Croatia, Cypress, Czech Republic, Denmark, Germany, Finland, France, Greenland, Hungary, Iceland, Italy, Luxembourg, Netherland, Norway, Sweden, Switzerland, Spain, Turkey, UK); (41 Rest of World countries - See Attachment for full country distribution + Japan, Portugal and Mexico)

Voluntary or Mandated:

Voluntary: Firm initiated