Roche Molecular Systems, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2551-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is intended for pipetting samples into a MagNA Pure 96 Processing Cartridge (purification rack position 1) before a purification run.
Product Classification:
Class II
Date Initiated: June 3, 2020
Date Posted: July 22, 2020
Recall Number: Z-2551-2020
Event ID: 85838
Reason for Recall:
When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.
Status: Terminated
Product Quantity: 917
Code Information:
All units when used with Sample Transfer protocol version 3.0
Distribution Pattern:
Worldwide distribution - US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated