Roche Molecular Systems, Inc.: Medical Device Recall in 2021 - (Recall #: Z-1413-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
cobas SARS-CoV-2 & Influenza A/B assay cobas Liat- Reagents, 2019-Novel Coronavirus Nucleic Acid for use on the cobas Liat System, Catalog Number: 09211101190
Product Classification:
Class II
Date Initiated: March 25, 2021
Date Posted: April 21, 2021
Recall Number: Z-1413-2021
Event ID: 87622
Reason for Recall:
Invalid or false positive results for one or more targets (SARS-CoV-2/Influenza A/Influenza B) using the cobas SARS-CoV-2 & Influenza A/B assay for use under emergency use authorization on the cobas Liat analyzer, showed abnormal PCR growth curves
Status: Ongoing
Product Quantity: US: 143,823 kits Ex-US: 57,219 kits
Code Information:
All lot numbers
Distribution Pattern:
Nationwide WorldWide
Voluntary or Mandated:
Voluntary: Firm initiated