Roche Molecular Systems, Inc.: Medical Device Recall in 2021 - (Recall #: Z-1414-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

cobas Liat Analyzer: respiratory virus panel nucleic acid assay system Catalog Number: 07341920190

Product Classification:

Class II

Date Initiated: March 25, 2021
Date Posted: April 21, 2021
Recall Number: Z-1414-2021
Event ID: 87622
Reason for Recall:

Invalid or false positive results for one or more targets (SARS-CoV-2/Influenza A/Influenza B) using the cobas SARS-CoV-2 & Influenza A/B assay for use under emergency use authorization on the cobas Liat analyzer, showed abnormal PCR growth curves

Status: Ongoing
Product Quantity: US: 5,430 units ; OUS 1528 units
Code Information:

All serial ID numbers

Distribution Pattern:

Nationwide WorldWide

Voluntary or Mandated:

Voluntary: Firm initiated