Roho Inc.: Medical Device Recall in 2017 - (Recall #: Z-1921-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology.

Product Classification:

Class II

Date Initiated: March 20, 2017
Date Posted: May 10, 2017
Recall Number: Z-1921-2017
Event ID: 76836
Reason for Recall:

When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under inflated Sensor Ready Cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.

Status: Terminated
Product Quantity: 5646 units
Code Information:

Serial Numbers: A2920163 through A3185287 Model Numbers: 1R1010H-CA-SC, 1R109M-CA-SR, 1R1211H-CH-SR, 1R88H-CA-SR, 1R1010H-CA-SCDM, 1R109M-CF-SR, 1R1212H-CA-SR, 1R88H-CF-SR, 1R1010H-CA-SR, 1R109M-CH-SR, 1R1212H-CH-SR, 1R88H-CH-SR, 1R1010H-CF-SCDM, 1R1110H-CA-SC, 1R129H-CA-SR, 1R88M-CA-SR, 1R1010H-CF-SR, 1R1110H-CA-SR, 1R1310H-CA-SR, 1R88M-CF-SR, 1R1010H-CH-SR, 1R1110H-CF-SR, 1R1310H-CF-SR, 1R89H-CA-SR, 1R1010M-CA-SC, 1R1110H-CH-SC, 1R1311H-CA-SR, 1R89H-CF-SR, 1R1010M-CA-SCDM, 1R1110H-CH-SR, 1R1311H-CH-SR, 1R89M-CA-SR, 1R1010M-CA-SR, 1R1110M-CA-SR, 1R1411H-CA-SR, 1R89M-CF-SR, 1R1010M-CF-SCDM, 1R1110M-CF-SR, 1R1411H-CF-SR, 1R910H-CA-SR, 1R1010M-CF-SR, 1R1110M-CH-SC, 1R1511H-CA-SR, 1R910H-CF-SR, 1R1010M-CH-SR, 1R1110M-CH-SR, 1R66H-CA-SR, 1R910H-CH-SC, 1R1011H-CA-SC, 1R1111H-CA-SC, 1R66H-CF-SR, 1R910H-CH-SR, 1R1011H-CA-SR, 1R1111H-CA-SR, 1R67H-CF-SR, 1R910M-CA-SR, 1R1011H-CF-SR, 1R1111H-CF-SR, 1R77H-CA-SR, 1R910M-CF-SR, 1R1011H-CH-SC, 1R1111H-CH-SC, 1R77H-CF-SR, 1R911H-CA-SC, 1R1011H-CH-SR, 1R1111H-CH-SR, 1R77M-CA-SR, 1R911H-CA-SR, 1R1011M-CA-SC, 1R1111M-CA-SR, 1R77M-CF-SR, 1R911H-CF-SR, 1R1011M-CA-SR, 1R1111M-CF-SR, 1R78H-CA-SR, 1R911H-CH-SR, 1R1011M-CF-SR, 1R1111M-CH-SR, 1R78H-CF-SR, 1R98H-CA-SR, 1R1012H-CA-SR, 1R1112H-CA-SR, 1R78M-CA-SR, 1R98H-CF-SR, 1R1012H-CF-SR, 1R1112H-CF-SR, 1R78M-CF-SR, 1R99H-CA-SC, 1R1012H-CH-SR, 1R1112H-CH-SR, 1R79H-CA-SR, 1R99H-CA-SCDM, 1R108H-CA-SR, 1R119H-CA-SC, 1R79H-CF-SR, 1R99H-CA-SR, 1R108H-CF-SR, 1R119H-CA-SR, 1R810H-CA-SR, 1R99H-CF-SCDM, 1R109H-CA-SCDM, 1R119H-CF-SR, 1R810H-CF-SR, 1R99H-CF-SR, 1R109H-CA-SR, 1R1210H-CA-SC, 1R810M-CA-SR, 1R99H-CH-SR, 1R109H-CF-SCDM, 1R1210H-CA-SR, 1R87H-CA-SC, 1R99M-CA-SR, 1R109H-CF-SR, 1R1210H-CF-SR, 1R87H-CA-SR, 1R99M-CF-SR, 1R109H-CH-SC, 1R1210H-CH-SR, 1R87H-CF-SR, 1R99M-CH-SR, 1R109H-CH-SR, 1R1211H-CA-SR

Distribution Pattern:

Nationwide, Sweden, Germany, New Zealand, Finland, Australia, Singapore, Korea, Luxembourg, Czech Republic, South Africa, Canada, Switzerland, United Arab Emirates, Spain, Austria, Hong Kong, Iceland, Japan, Canada, Italy, Ireland, France, United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated