S.A.E. AFIKIM: Medical Device Recall in 2013 - (Recall #: Z-0115-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Breeze C Scooter is a motorized electric scooter for elderly people. Brand name Breeze, model C. It is intended to provide increased mobility for elderly and/or disabled person and it is also an aid to independent living. Products can be identified by their color; the recalled products are either metallic blue or metallic silver. The products have two rear wheels and have either one or two wheels in the front. The scooters have an adjustable swivel seat and a front storage basket; some scooters may be fitted with an optional larger rear basket.

Product Classification:

Class II

Date Initiated: September 27, 2013
Date Posted: November 6, 2013
Recall Number: Z-0115-2014
Event ID: 66410
Reason for Recall:

Breeze C Scooter, motorized 3-wheeled electric scooter for elderly people, is recalled because it may become unresponsive during operation due to an electrical contact which may fault due to extensive wear, which potentially could result in the driver not being able to stop the scooter during operation except by turning off the ignition key.

Status: Terminated
Product Quantity: 69 units
Code Information:

Serial Numbers: 10390024, 10390025, 10390029, 10390030, 10390034, 10390037, 10490100, 10490108, 10490111, 10490113, 10490119, 10490121, 10490124, 11390058, 11390060, 11390061, 11390069, 11390074, 11390095, 11390101, 11490111, 11490147, 11490176, 11490189, 11490277, 11490304, 11490305, 11490425, 11490426, 11490433, 11490435, 11490445, 12390096, 12390097, 12390098, 12390099, 12390101, 12390102, 12390103, 12390104, 12390105, 12390106, 12390107, 12390117, 12390120, 12490169, 12490170, 12490172, 12490173, 12490174, 12490175, 12490176, 12490177, 12490178, 12490179, 12490180, 12490181, 12490182, 12490183, 12490184, 12490185, 12490186, 12490187, 12490188, 12490189, 12490190, 12490191, 12490192, and 12490193.

Distribution Pattern:

Nationwide Distribution: US including the states of California, Florida, Georgia, Illinois, Maine, Maryland, Massachusetts, Michigan, Minnesota, New Jersey, New York, Ohio, Oklahoma, Pennsylvania, Texas, Virginia, and Wisconsin.

Voluntary or Mandated:

Voluntary: Firm initiated