Sarnova HC, Llc: Medical Device Recall in 2024 - (Recall #: Z-0721-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Curaplex Cricothyrotomy Field Kit with Instruments, Kwik Cric MVP, 6.0mm, Item Number 351631i-A. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.
Product Classification:
Class II
Date Initiated: November 6, 2023
Date Posted: January 24, 2024
Recall Number: Z-0721-2024
Event ID: 93665
Reason for Recall:
It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.
Status: Ongoing
Product Quantity: 15 kits
Code Information:
Lot Numbers: ASM0037685 ASM0037887
Distribution Pattern:
US Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated