Schaerer Medical USA Inc: Medical Device Recall in 2016 - (Recall #: Z-0371-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

The surgical table is a stand alone unit. Three (3) floor locking cylinders are attached to the base of the surgical table. The individual cylinders are replaceable. Product Usage: The cylinders are used to lock the surgical table in place.

Product Classification:

Class II

Date Initiated: September 27, 2016
Date Posted: November 16, 2016
Recall Number: Z-0371-2017
Event ID: 75288
Reason for Recall:

The floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect "O" ring which may cause the cylinder to leak.

Status: Terminated
Product Quantity: 4 surgical tables with defective floor locking cylinders were distributed. Each table contains 3 cylinders for a total of 12 cylinders.
Code Information:

Lot Codes for the four (4) model 7300 Surgical Tables TFX-1653, TFX1654, TFX1655, TFX1656

Distribution Pattern:

Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed

Voluntary or Mandated:

Voluntary: Firm initiated