Product Description:

Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders are part number 85118. Product Usage: The cylinders are used to lock the surgical table in place.

Product Classification:

Class II

Date Initiated: September 27, 2016
Date Posted: November 16, 2016
Recall Number: Z-0372-2017
Event ID: 75288
Reason for Recall:

The floor locking cylinders are contained in the base of the Schaerer 7300 model surgical tables. The cylinders were manufactured with the incorrect "O" ring which may cause the cylinder to leak.

Status: Terminated
Product Quantity: 10 replacement cylinders.
Code Information:

Lot code of the replacement cylinders, 2016-02-01

Distribution Pattern:

Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed

Voluntary or Mandated:

Voluntary: Firm initiated