Seca: Medical Device Recall in 2020 - (Recall #: Z-1429-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009
Class II
It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.
serial numbers: 6002142000002 - 6002142001101
US: CA, MI, WA, WV, VA, NM, LA, TN, IL, PA, OH, FL, WA, OR, NY, CO, IA, MN, TX, MA, NC, LA, KY, NM AL,DC, NJ, AK, AZ, NV, SC, HI, MD, WI, RI, DE, MO OUS: Canada
Voluntary: Firm initiated