Seca: Medical Device Recall in 2020 - (Recall #: Z-1429-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009

Product Classification:

Class II

Date Initiated: June 19, 2019
Date Posted: March 11, 2020
Recall Number: Z-1429-2020
Event ID: 84817
Reason for Recall:

It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.

Status: Ongoing
Product Quantity: 254 scale carts
Code Information:

serial numbers: 6002142000002 - 6002142001101

Distribution Pattern:

US: CA, MI, WA, WV, VA, NM, LA, TN, IL, PA, OH, FL, WA, OR, NY, CO, IA, MN, TX, MA, NC, LA, KY, NM AL,DC, NJ, AK, AZ, NV, SC, HI, MD, WI, RI, DE, MO OUS: Canada

Voluntary or Mandated:

Voluntary: Firm initiated