Sentec AG: Medical Device Recall in 2012 - (Recall #: Z-0433-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

SenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The SenTec V-Sign Membrane Changer is indicated for use with the V-Sign Sensor only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients.

Product Classification:

Class II

Date Initiated: November 5, 2012
Date Posted: November 28, 2012
Recall Number: Z-0433-2013
Event ID: 63631
Reason for Recall:

A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

Status: Terminated
Product Quantity: 195 boxes (9 per box)
Code Information:

Lot numbers: 1CF5, 1D21, 1D60, 1DC6, 1DED, 1E0D

Distribution Pattern:

Distributed only in MO.

Voluntary or Mandated:

Voluntary: Firm initiated