Shanghai United Imaging Healthcare Co., Ltd.: Medical Device Recall in 2020 - (Recall #: Z-2522-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
uCT 760 System, Computed Tomography X-ray System
Class II
Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.
Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with software version R001.11.0.785930-Re-1150.
Distributed to two customers. Both customers located in Houston, Texas.
Voluntary: Firm initiated