SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION: Medical Device Recall in 2024 - (Recall #: Z-1292-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Shimadzu, Trinias, Digital Angiography System

Product Classification:

Class II

Date Initiated: January 29, 2024
Date Posted: March 20, 2024
Recall Number: Z-1292-2024
Event ID: 93933
Reason for Recall:

Due to capacitor manufacturing issue, their is a potential that Flat Panel Device may have abnormal display or no function.

Status: Ongoing
Product Quantity: 1
Code Information:

Serial Number: 41EFFF7D3001/UDI: (01)04540217058020

Distribution Pattern:

US Nationwide distribution in the states of OH.

Voluntary or Mandated:

Voluntary: Firm initiated