Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2012 - (Recall #: Z-2172-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

EasyLink(TM) Informatics System, Part Number 1000034805. EasyLink(TM) Informatics System is classified as a data calculator and processing module for use with In-Vitro diagnostics devices.

Product Classification:

Class II

Date Initiated: June 7, 2012
Date Posted: August 15, 2012
Recall Number: Z-2172-2012
Event ID: 62448
Reason for Recall:

There is a potential for the EasyLink system to display and transmit to the Laboratory Information System networks an incorrect result following the EasyLink systems application of the configured result number format. The problem is encountered when the result number format configured on the Sysmex CA-7000 or CA-1500 coagulation analyzer does not match the EasyLink system settings on the following

Status: Terminated
Product Quantity: 237
Code Information:

Version 5.0 and Version 5.0 Service Pack 1 through 4

Distribution Pattern:

Worldwide distribution, including the countries of Australia, Austria, Belgium, Canada, Czech Republic, France, Germany, Italy, Netherlands, New Zealand, Slovakia, Spain, Switzerland and the United States, with nationwide distribution. Center Recall Depth is Retail.

Voluntary or Mandated:

Voluntary: Firm initiated