Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0664-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). Homogeneous enzyme immunoassay intended for use in the quantitative analysis of Phenobarbital in human serum or plasma.

Product Classification:

Class II

Date Initiated: December 10, 2012
Date Posted: January 16, 2013
Recall Number: Z-0664-2013
Event ID: 63892
Reason for Recall:

Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R) Emit(R) 2000 Phenobarbital (4D019UL) lots E1, E2, and E3 as an application if the AU(R) System is used, the issue may be experienced. Firm has found that within-run imprecision of the recalled lots frequency of outliers that differ from the expected value by 20%. This is an expansion of

Status: Terminated
Product Quantity: 1979 cartons
Code Information:

Emit(R) 2000 Phenobarbital Assay (OSR4D229) lots 1237 and 1264 and/or Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots E1, E2, and E3.

Distribution Pattern:

Distributed in the states of CA, FL, IL, KS, MA, MI, MN, MS, NM, NY, NC, OH, OR, PA, TN, TX, UT, and WA.

Voluntary or Mandated:

Voluntary: Firm initiated