Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0745-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.
Class III
Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.
Siemens material number 10445010 -- and lot numbers 2DN060, exp 2013-10-01; 2FN063, exp 2013-12-01; 2GN068, exp 2014-01-01; and 2JN073, exp 2014-03-01.
Nationwide Distribution.
Voluntary: Firm initiated