Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1049-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Siemens Dimension(R) Phosphorus (PHOS) Flex(R) reagent cartridge (DF61) The Phosphorus (PHOS) method used on the Dimension(R) clinical chemistry system is an in vitro diagnostic test intended for the quantitative measurement of phosphorus in human serum, plasma, and urine.

Product Classification:

Class II

Date Initiated: February 22, 2013
Date Posted: April 10, 2013
Recall Number: Z-1049-2013
Event ID: 64463
Reason for Recall:

Siemens Healthcare Diagnostics has initiated a voluntary field action for Dimension(R) PHOS Flex(R) Reagent due to reagent interaction causing falsely depressed Creatinine results when EZCR is processed from open wells that are in close proximity to open wells of PHOS reagent.

Status: Terminated
Product Quantity: 60,674
Code Information:

Lot numbers DC3078, exp 3/19/13; DA3092, exp 4/2/13; FB3109, exp 4/19/13; BC3127, exp 5/7/13; BB3152, exp 6/1/13; EC3166, exp 6/15/13; BB3190, exp 7/9/13; DA3212, exp 7/31/13; EA3233, exp 8/21/13; FB3260, exp 9/17/13; DA3319, exp 11/15/13; DA3320, exp 11/16/13; GC3337, exp 12/3/13; FB3346, exp 12/12/13; CB4014, exp 1/14/14; and FB4029, exp 1/29/14.

Distribution Pattern:

Nationwide Distribution including the states of AK, AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.

Voluntary or Mandated:

Voluntary: Firm initiated