Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1073-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination of the activated partial thromboplastin time (APTT).

Product Classification:

Class II

Date Initiated: March 11, 2013
Date Posted: April 17, 2013
Recall Number: Z-1073-2013
Event ID: 63804
Reason for Recall:

Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.

Status: Terminated
Product Quantity: 13,083
Code Information:

Catalog numbers B4219-1 and B4219-2: lot numbers 547311, exp 3/22/2013; 547312, exp 3/22/2013 and 547316A, exp 4/28/2013.

Distribution Pattern:

Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV and Puerto Rico and the country of Canada.

Voluntary or Mandated:

Voluntary: Firm initiated