Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2013 - (Recall #: Z-2120-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Microscan Synergies Plus Negative Combo 2 Product Usage: MicroScan¿ Synergies plus¿ Panels & MicroScan¿ Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.

Product Classification:

Class I

Date Initiated: August 21, 2013
Date Posted: September 18, 2013
Recall Number: Z-2120-2013
Event ID: 66058
Reason for Recall:

Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.

Status: Terminated
Product Quantity: 2158 boxes, 43,160 panels
Code Information:

Internal Number - B1025-108; Catalog Number - 10444747; Lot code/expiration date: 2012-06-27, 2012-06-27; 2012-08-10, 2012-08-10; 2012-10-12, 2012-10-12; 2012-11-30, 2012-11-30; 2013-01-06, 2013-01-06; 2013-02-01, 2013-02-01; 2013-03-27, 2013-03-27; 2013-10-02, 2013-10-02; 2013-11-21, 2013-11-21; 2014-01-24, 2014-01-24; 2014-03-28, 2014-03-28; 2014-05-03, 2014-05-03.

Distribution Pattern:

Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, South Africa, Kazakhstan, Belgium, Great Britain, Turkey, Spain, Italy, Saudi Arabia, Qutar, Poland, Iraq, Unite Arab Emrites, Algeria, Oman, Greece, Portugal, and Ireland.

Voluntary or Mandated:

Voluntary: Firm initiated