Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0625-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

HB1C Flex(R) Reagent Cartridge (DF105A, Siemens Material Number 10483822) for the Dimension(R) Clinical Systems. Dimension(R) Hemoglobin A1c (HB1C) Flex(R) reagent Cartridge on the Dimension(R) clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus.

Product Classification:

Class II

Date Initiated: April 22, 2013
Date Posted: January 15, 2014
Recall Number: Z-0625-2014
Event ID: 65409
Reason for Recall:

Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.

Status: Terminated
Product Quantity: 11006 Total
Code Information:

Catalog number DF105A (Siemens Material Number 10483822) - Lot numbers GA3099, exp. 4/9/13; GA3141, exp. 5/21/13; GA3197, exp. 7/16/13; GA3247, exp. 9/4/13; GA3113, exp. 4/23/13; GA3162, exp. 6/11/13; GB3211, exp. 7/30/13; GA3267, exp. 9/24/13; GA3134, exp. 5/14/13; GA3169, exp. 6/18/13; GA3232, exp. 8/20/13; GB3281, exp. 10/8/13, GC3302, exp. 10/29/13.

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and internationally to Canada and Mexico.

Voluntary or Mandated:

Voluntary: Firm initiated