Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0625-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
HB1C Flex(R) Reagent Cartridge (DF105A, Siemens Material Number 10483822) for the Dimension(R) Clinical Systems. Dimension(R) Hemoglobin A1c (HB1C) Flex(R) reagent Cartridge on the Dimension(R) clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus.
Class II
Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.
Catalog number DF105A (Siemens Material Number 10483822) - Lot numbers GA3099, exp. 4/9/13; GA3141, exp. 5/21/13; GA3197, exp. 7/16/13; GA3247, exp. 9/4/13; GA3113, exp. 4/23/13; GA3162, exp. 6/11/13; GB3211, exp. 7/30/13; GA3267, exp. 9/24/13; GA3134, exp. 5/14/13; GA3169, exp. 6/18/13; GA3232, exp. 8/20/13; GB3281, exp. 10/8/13, GC3302, exp. 10/29/13.
Worldwide Distribution - USA (nationwide) and internationally to Canada and Mexico.
Voluntary: Firm initiated