Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2015 - (Recall #: Z-0340-2016)
See the recall detail below. You can also see other recalls from the same firm in 2015.
ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems. For in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems.
Class II
Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.
10379673 (160 tests kit), 10485591 (800 tests kit) Lot 230 (SMN 10485591 and 10379673) and lot 231(SMN 10485591)
Distributed in the states of AZ, NY, CA, IN, LA, MN, OH, MI, NM, and GA, and the countries of Mexico and Canada.
Voluntary: Firm initiated