Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2015 - (Recall #: Z-0468-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualitative and Quantitative Factor Deficiency

Product Classification:

Class II

Date Initiated: November 19, 2015
Date Posted: December 30, 2015
Recall Number: Z-0468-2016
Event ID: 72752
Reason for Recall:

Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Systems. This could lead, under certain circumstances, to an erroneously increased result with a Factor VIII Chromogenic Assay.

Status: Terminated
Product Quantity: Domestic: 52 kits, Foreign: 623 kits
Code Information:

Lot Number 44637, exp. 2016-11-30

Distribution Pattern:

Worldwide Distribution - US Distribution to the state of : Massachusetts and to the countries of : Austria, Czech Republic, France, Germany, Hungary, Italy, Poland, Serbia, Spain and Sweden.

Voluntary or Mandated:

Voluntary: Firm initiated