Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1399-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

ADVIA Chemistry XPT System, Siemens Material Number 10723034, Software Version 1.0.2, Siemens Material Number 11219493, IVD. The ADVIA XPT Chemistry System is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour.

Product Classification:

Class II

Date Initiated: February 11, 2015
Date Posted: April 15, 2015
Recall Number: Z-1399-2015
Event ID: 70591
Reason for Recall:

Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The system may cause samples to remain in an Inprocess state. Test results on a sample that is held Inprocess will not transmit to the LIS. Manual intervention is necessary to complete the processing of the samples that are held Inprocess.

Status: Terminated
Product Quantity: 16 units: 3 Domestic, 13 Foreign
Code Information:

Software Version 1.0.2; Lot Numbers: CA1275000080008, CA1275000090009, CA1275000060006, CA1275000050005, CA1275000070007, CA1275000130013, CA1275000100010, CA1275000120012, CA1275000170017, CA1275000260026, CA1275000270027, CA1275000280028, CA1275000240024, CA1275000200020, CA1275000220022, CA1275000410041

Distribution Pattern:

Distributed in the states of AZ, CA, and WA, and the countries of Germany, Italy, Spain, and UK.

Voluntary or Mandated:

Voluntary: Firm initiated