Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1457-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.
Class II
Siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier results when running with Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents.
Catalog Number: B01-4840-01; Material Number (SMN)10311896, Lot Numbers: 318706, 332463, expiration dates 2015-06-30, 2015-08-31
Worldwide Distribution - US (nationwide) Distribution and to the countries of : Thailand, Pakistan, Vatican, Guadeloupe, Reunion, French Polynesia , Serbia, Martinique, India, China, Singapore, Brazil, South Africa, Mexico, Canada, Israel, Australia, Egypt, Russian Federation, Indonesia, Rep. Korea (S) ,Unit.Arab Emir., Taiwan ,Argentina ,Chile, Malaysia ,Colombia, Peru, Kazakhstan, Paraguay, Vietnam, Algeria, Austria, Belgium, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Greece, Hungary, Italy, Lithuania, Luxembourg, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Slovakia and Turkey.
Voluntary: Firm initiated