Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2015 - (Recall #: Z-2054-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1. In vitro diagnostic analyzer.

Product Classification:

Class II

Date Initiated: May 21, 2015
Date Posted: July 15, 2015
Recall Number: Z-2054-2015
Event ID: 71315
Reason for Recall:

Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification. Issue #2: Timing issue in Vista 1500 may cause wrong reagent or no reagent delivery. This may lead to unflagged, unexpected low results and results flagged with assay errors.

Status: Terminated
Product Quantity: 2315
Code Information:

Siemens Material Number (SMN): 10284473, 10488224, 10444801, 10444802

Distribution Pattern:

Worldwide Distribution - Nationwide Distribution including Puerto Rico., and the countries of : Austria, Australia, Belgium, Canada, Denmark, France, Finland, Germany, Japan, Italy, New Zealand, Netherlands, Czech Republic, Portugal, Republic Korea, Norway, Spain, Switzerland, Slovakia, Saudi Arabia, Slovenia, and United Kingdom,

Voluntary or Mandated:

Voluntary: Firm initiated