Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2016 - (Recall #: Z-0082-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

SIEMENS IMMULITE/IMMULITE 2000 Systems Rubella IgM (RUM), REF/Catalog Number for US: L2KRM2(D), Siemens Material Number (SMN) for US: 10374021, REF/Catalog # for OUS: L2KRM2, SMN for OUS: 10381327; Unique Device Identification Number (UDI) 00630414961996; IVD --- For in vitro diagnostic use with the IMMULITE/IMMULITE 1000/IMMULITE 2000/IMMULITE 2000 XPi Systems Analyzers-for the qualitative detection of IgM antibodies to Rubella virus in human serum or plasma (EDTA and heparinized), as an aid in the presumptive diagnosis of an acute or recent rubella infection, particularly in women of childbearing age

Product Classification:

Class II

Date Initiated: August 24, 2016
Date Posted: October 19, 2016
Recall Number: Z-0082-2017
Event ID: 75037
Reason for Recall:

Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.

Status: Terminated
Product Quantity: Domestic: 177 Kits; Foreign: 3435 Kits
Code Information:

Kit lots 236M, 237, 238, 239, 239L (OUS) and D236M and D238 (US)

Distribution Pattern:

Worldwide Distribution n- US (Nationwide) and Foreign distribution to the following countries: Argentina, Bangladesh, Brazil, Bulgaria, China, Colombia, Costa Rica, Cyprus, Egypt, France, Germany, Greece, Guatemala, India, Indonesia, Israel, Italy, Kazakhstan, Macedonia, Mexico, Paraguay, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Africa, Syria, United Arab Emirates, and Uruguay.

Voluntary or Mandated:

Voluntary: Firm initiated