Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2016 - (Recall #: Z-1618-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of methotrexate in human serum or plasma.

Product Classification:

Class II

Date Initiated: March 8, 2016
Date Posted: May 18, 2016
Recall Number: Z-1618-2016
Event ID: 73594
Reason for Recall:

Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability claim of 20 hours. Testing using reagent at the end of the 20 hours On-Board Stability has shown that values of methotrexate for patient and quality control samples increased.

Status: Terminated
Product Quantity: 18364 units
Code Information:

Lot # F1, F2, G1, G2, H1, and H2

Distribution Pattern:

Units were shipped to MI, FL, NJ, NY, DC, CA, GA, MI, MT, VA, NC, TX, PA, Australia, Belgium, Czech Republic, Finland, Japan, Republic of Korea, Italy, Saudi Arabia, Slovakia, Slovenia, Spain and Switzerland.

Voluntary or Mandated:

Voluntary: Firm initiated