Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2016 - (Recall #: Z-1622-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Dimension Vista Blood Urea Nitrogen (BUN) Flex reagent cartridge The BUN method is an in vitro diagnostic test for the quantitative measurement of urea nitrogen (an end product of nitrogen metabolism) in human serum, plasma, and urine on the Dimension Vista System.

Product Classification:

Class II

Date Initiated: February 16, 2016
Date Posted: May 18, 2016
Recall Number: Z-1622-2016
Event ID: 73592
Reason for Recall:

Siemens Healthcare Diagnostics has determined that Dimension Vista Blood Urea Nitrogen (BUN) lots 15215AE, 15243BB, 15264BA, 15299BB, 15300BA and 15341AC may exhibit inaccurate patient and/or Quality Control results. The issue has been isolated to Flex reagen cartridges manufactured from specific mold cavities.

Status: Terminated
Product Quantity: 18668 units
Code Information:

Device Listing # D011374, Lot # 15341AC, 15320BB, 15300BA, 15299BB, 15264BA, 15243BB and 15215AE

Distribution Pattern:

Worldwide distribution - US (nationwide) distribution including Puerto Rico and to the states of : AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. , and to the countries of : Canada, United Arab Emir, Australia, Austria, Belgium, Switzerland, Bahamas, Czech Republic, Germany, Denmark, Spain, Finland, France, Italy, Netherland, Norway, Portugal, Slovakia, Slovenia, Japan, Republic of Korea, New Zealand.

Voluntary or Mandated:

Voluntary: Firm initiated