Siemens Healthcare Diagnostics, Inc.: Medical Device Recall in 2016 - (Recall #: Z-1622-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
Dimension Vista Blood Urea Nitrogen (BUN) Flex reagent cartridge The BUN method is an in vitro diagnostic test for the quantitative measurement of urea nitrogen (an end product of nitrogen metabolism) in human serum, plasma, and urine on the Dimension Vista System.
Class II
Siemens Healthcare Diagnostics has determined that Dimension Vista Blood Urea Nitrogen (BUN) lots 15215AE, 15243BB, 15264BA, 15299BB, 15300BA and 15341AC may exhibit inaccurate patient and/or Quality Control results. The issue has been isolated to Flex reagen cartridges manufactured from specific mold cavities.
Device Listing # D011374, Lot # 15341AC, 15320BB, 15300BA, 15299BB, 15264BA, 15243BB and 15215AE
Worldwide distribution - US (nationwide) distribution including Puerto Rico and to the states of : AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. , and to the countries of : Canada, United Arab Emir, Australia, Austria, Belgium, Switzerland, Bahamas, Czech Republic, Germany, Denmark, Spain, Finland, France, Italy, Netherland, Norway, Portugal, Slovakia, Slovenia, Japan, Republic of Korea, New Zealand.
Voluntary: Firm initiated